Instruction ManualNo. 012-09316CDataStudio New Features GuideInstallationOn a computer running Windows XP, Windows Vista, Windows 7, or Mac OS X, you
DataStudio New Features Calibration10®One-point (Adjust Offset Only)In a one-point offset calibration, you reset only one point. The line shifts so th
Model No. CI-6870 Calibration11®Enter knownvalues hereRawCalibratedobvious meaning. Depending on which sensor you are using, the raw measurement may
DataStudio New Features Calibration12®Enter knownvalue hereRawCalibratedOne-Point (Adjust Slope Only) Calibration ProcedureA one-point slope calibrati
Model No. CI-6870 Other Information13®Click toautomaticallyenterStandardValue andSensor ValueRawCalibrated5. Place all of the probes so that they are
DataStudio New Features Installation2®I. Software Installation1. Log onto the computer with an administrator account.2. Insert the DataStudio CD. In W
Model No. CI-6870 Installation3®3. If you installed WavePort, click WavePort in the Product list to select it, then do one of the following.• If you
DataStudio New Features Installation4®III. Additional RecommendationsInstall QuickTimeDataStudio’s movie playback feature requires the free QuickTime
Model No. CI-6870 Installation5®Run the DataStudio TutorialThe DataStudio Tutorial included on the DataStudio installer CD is an interactive multimed
DataStudio New Features Experiment Setup Window6®Experiment Setup WindowTo support PASCO’s growing range of interfaces and sensors, DataStudio 1.9.6 f
Model No. CI-6870 Experiment Setup Window7®Click a port on the picture of an interface to manually add a sensor. Press the Back-space or Delete key o
DataStudio New Features Calibration8®If you are using a ScienceWorkshop interface, the Timer Setup dialog box will open. For instructions on creating
Model No. CI-6870 Calibration9®This function can be graphically represented as a line (right). Two points, Point 1 and Point 2, define the line. In t