Instruction Manual
012-13883A
Advanced Chemistry Sensor
PS-2172
1
Included Equipment Part Number
1. Advanced Chemistry Sensor PS-2172
2. Stainless Steel Temperature Probe PS-2153
3. Conductivity Probe (10x) PS-2571
4. pH Electrode PS-2573
5. Polyurethane Tubing (60 cm long, inch ID)
6. Quick-release Connectors (qty. 4)
7. Tubing Connectors (qty. 4)
8. Syringe (60 cc)
Required Equipment
PASPORT Interface See PASCO catalog or www.pasco.com
Optional Equipment
Fast-response Temperature Probe PS-2135 (3-pack)
Skin/surface Temperature Probe PS-2131
Various Ion-selective Electrodes (ISE) See PASCO catalog or www.pasco.com
Oxidation Reduction Potential (ORP) Electrode CI-6716
1
8
Page view 0
1 2 3 4 5 6 7 8 9 10

Summary of Contents

Page 1 - Advanced Chemistry Sensor

®Instruction Manual012-13883AAdvanced Chemistry SensorPS-21721Included Equipment Part Number1. Advanced Chemistry Sensor PS-21722. Stainless Steel Tem

Page 2 - Introduction

®Advanced Chemistry Sensor Technical Support10Technical SupportFor assistance with any PASCO product, contact PASCO at:Limited WarrantyFor a descripti

Page 3 - Temperature

®Advanced Chemistry Sensor Introduction2IntroductionThe PS-2172 Advanced Chemistry Sensor combines four sensors in a single unit:• Temperature• pH, OR

Page 4 - Measuring pH

®Model No. PS-2172 Temperature3SPARKvue/SPARK SLS, refer to the SPARKvue/SPARK SLS Online Help or User’s Guide.Xplorer GLX SetupIf you are using an X

Page 5 - Conductivity

®Advanced Chemistry Sensor pH/ISE/ORP4Measuring pH The pH electrode produces a voltage proportional to the pH of the solution that it is immersed in.

Page 6 - Conductivity Calibration

®Model No. PS-2172 Pressure5PressureThis component of the multi-sensor measures absolute gas pressure from 20 kPa to 400 kPa. The measurement can be

Page 7 - Specifications

®Advanced Chemistry Sensor Conductivity6store. The table (right) lists typical conductivity values for common aqueous solutions at 25 °C.As a rule of

Page 8

®Model No. PS-2172 Specifications7SpecificationsGeneralComponent Sensors Temperature, pH/ISE/ORP probe voltage, Absolute Pressure, VoltageMax. Sampli

Page 9

®Advanced Chemistry Sensor Specifications8Appendix A: DataStudio CalibrationIn DataStudio, click the Setup button to open the Experiment Setup window;

Page 10 - Technical Support

®Model No. PS-2172 Specifications9Appendix B: Reference Solutions for Conductivity CalibrationReference SolutionApproximation Normality of SolutionMe